mirror of
https://gitlab.com/pholy/OSCAR-code.git
synced 2025-04-05 02:30:44 +00:00
Exclude additional invalid samples in PRS1 oximetry.
Also clean up some function names and remove unnecessary code.
This commit is contained in:
parent
5f0960aa4d
commit
01c7f7cdc0
@ -29,6 +29,7 @@
|
||||
<li>[fix] Fix the pressure waveform scale for the BiPAP autoSV Advanced 30 (960T)</li>
|
||||
<li>[fix] Add support for rise time mode on DreamStation BiPAP devices (600X-700X).</li>
|
||||
<li>[fix] Remove the ramp time and pressure settings when the ramp is disabled on pre-DreamStation devices.</li>
|
||||
<li>[fix] Improve import of Philips Respironics oximetry data.</li>
|
||||
<li>[fix] Fix occasional failure to save imported Viatom data.</li>
|
||||
</ul>
|
||||
<p>
|
||||
|
@ -8149,7 +8149,7 @@ QList<PRS1DataChunk *> PRS1Import::CoalesceWaveformChunks(QList<PRS1DataChunk *>
|
||||
}
|
||||
|
||||
|
||||
void PRS1Import::ParseOximetry()
|
||||
void PRS1Import::ImportOximetry()
|
||||
{
|
||||
int size = oximetry.size();
|
||||
|
||||
@ -8205,10 +8205,10 @@ void PRS1Import::ImportOximetryChannel(ChannelID channel, QByteArray & data, qui
|
||||
quint64 start_ti;
|
||||
int start_i;
|
||||
|
||||
// Split eventlist on invalid values (255)
|
||||
// Split eventlist on invalid values (254-255)
|
||||
for (int i=0; i < data.size(); i++) {
|
||||
unsigned char value = raw[i];
|
||||
bool valid = (value != 255);
|
||||
bool valid = (value < 254);
|
||||
|
||||
if (valid) {
|
||||
if (pending_samples == false) {
|
||||
@ -8218,7 +8218,7 @@ void PRS1Import::ImportOximetryChannel(ChannelID channel, QByteArray & data, qui
|
||||
}
|
||||
|
||||
if (channel == OXI_Pulse) {
|
||||
if (value > 200) UNEXPECTED_VALUE(value, "<= 200 bpm");
|
||||
if (value > 240) UNEXPECTED_VALUE(value, "<= 240 bpm");
|
||||
} else {
|
||||
if (value > 100) UNEXPECTED_VALUE(value, "<= 100%");
|
||||
}
|
||||
@ -8242,7 +8242,7 @@ void PRS1Import::ImportOximetryChannel(ChannelID channel, QByteArray & data, qui
|
||||
}
|
||||
|
||||
|
||||
void PRS1Import::ParseWaveforms()
|
||||
void PRS1Import::ImportWaveforms()
|
||||
{
|
||||
int size = waveforms.size();
|
||||
quint64 s1, s2;
|
||||
@ -8429,25 +8429,16 @@ bool PRS1Import::ParseSession(void)
|
||||
// Parse .005 Waveform files
|
||||
waveforms = ReadWaveformData(m_wavefiles, "Waveform");
|
||||
|
||||
if (session->eventlist.contains(CPAP_FlowRate)) {
|
||||
if (waveforms.size() > 0) {
|
||||
// Delete anything called "Flow rate" picked up in the events file if high-resolution data is present
|
||||
// TODO: Is this still used anywhere?
|
||||
qWarning() << session->session() << "Deleting flow rate events due to flow rate waveform data";
|
||||
session->destroyEvent(CPAP_FlowRate);
|
||||
}
|
||||
}
|
||||
|
||||
// Extract raw data into channels.
|
||||
ParseWaveforms();
|
||||
// Extract and import raw data into channels.
|
||||
ImportWaveforms();
|
||||
}
|
||||
|
||||
if (!m_oxifiles.isEmpty()) {
|
||||
// Parse .006 Waveform files
|
||||
oximetry = ReadWaveformData(m_oxifiles, "Oximetry");
|
||||
|
||||
// Extract raw data into channels.
|
||||
ParseOximetry();
|
||||
// Extract and import raw data into channels.
|
||||
ImportOximetry();
|
||||
}
|
||||
|
||||
save = true;
|
||||
|
@ -313,10 +313,10 @@ public:
|
||||
QList<PRS1DataChunk *> CoalesceWaveformChunks(QList<PRS1DataChunk *> & allchunks);
|
||||
|
||||
//! \brief Takes the parsed list of Flow/MaskPressure waveform chunks and adds them to the database
|
||||
void ParseWaveforms();
|
||||
void ImportWaveforms();
|
||||
|
||||
//! \brief Takes the parsed list of oximeter waveform chunks and adds them to the database.
|
||||
void ParseOximetry();
|
||||
void ImportOximetry();
|
||||
|
||||
//! \brief Adds a single channel of continuous oximetry data to the database, splitting on any missing samples.
|
||||
void ImportOximetryChannel(ChannelID channel, QByteArray & data, quint64 ti, qint64 dur);
|
||||
|
Loading…