Clean up remaining F3V6 import messages.
This commit is contained in:
parent
5835e6de9c
commit
46a077cb43
@ -3736,7 +3736,7 @@ bool PRS1DataChunk::ParseSummaryF3V6(void)
|
||||
//CHECK_VALUE(data[pos+4], 0x00); // 16-bit minutes in LL
|
||||
CHECK_VALUE(data[pos+5], 0x00);
|
||||
//CHECK_VALUE(data[pos+6], 0x0A); // 16-bit VS count
|
||||
CHECK_VALUE(data[pos+7], 0x00);
|
||||
//CHECK_VALUE(data[pos+7], 0x00); // We've actually seen someone with more than 255 VS in a night!
|
||||
//CHECK_VALUE(data[pos+8], 0x01); // 16-bit H count (partial)
|
||||
CHECK_VALUE(data[pos+9], 0x00);
|
||||
//CHECK_VALUE(data[pos+0xa], 0x00); // 16-bit H count (partial)
|
||||
@ -3824,11 +3824,11 @@ bool PRS1DataChunk::ParseSettingsF3V6(const unsigned char* data, int size)
|
||||
switch (code) {
|
||||
case 0: // Device Mode
|
||||
CHECK_VALUE(pos, 2); // always first?
|
||||
// TODO: We may need additional enums for these modes, the below are just a rough guess mapping for now.
|
||||
// TODO: We probably need additional enums for these modes, the below are just a rough guess mapping for now.
|
||||
switch (data[pos]) {
|
||||
case 1: cpapmode = MODE_BILEVEL_FIXED; break; // TODO This is marked "S - Bi-Flex" on reports.
|
||||
case 2: cpapmode = MODE_ASV; break; // TODO: This is marked as "S/T" on reports, is that spontaneous/timed? Pressure also seems variable!
|
||||
case 4: cpapmode = MODE_AVAPS; break; // "PC - AVAPS" on reports
|
||||
case 1: cpapmode = MODE_BILEVEL_FIXED; break; // "S" mode
|
||||
case 2: cpapmode = MODE_ASV; break; // "S/T" mode; pressure seems variable?
|
||||
case 4: cpapmode = MODE_AVAPS; break; // "PC" mode? Usually "PC - AVAPS", see setting 1 below
|
||||
default:
|
||||
UNEXPECTED_VALUE(data[pos], "known device mode");
|
||||
break;
|
||||
@ -3836,10 +3836,9 @@ bool PRS1DataChunk::ParseSettingsF3V6(const unsigned char* data, int size)
|
||||
this->AddEvent(new PRS1ParsedSettingEvent(PRS1_SETTING_CPAP_MODE, (int) cpapmode));
|
||||
break;
|
||||
case 1: // ???
|
||||
if (cpapmode == MODE_AVAPS) {
|
||||
CHECK_VALUE(data[pos], 2); // 2 when in AVAPS
|
||||
} else {
|
||||
CHECK_VALUES(data[pos], 0, 1); // 1 when in S - Bi-Flex, 0 when in S/T
|
||||
// How do these interact with the mode above?
|
||||
if (data[pos] != 2) { // 2 = AVAPS: usually "PC - AVAPS", sometimes "S/T - AVAPS"
|
||||
CHECK_VALUES(data[pos], 0, 1); // 0 = None, 1 = Bi-Flex
|
||||
}
|
||||
break;
|
||||
case 2: // ???
|
||||
@ -3864,7 +3863,8 @@ bool PRS1DataChunk::ParseSettingsF3V6(const unsigned char* data, int size)
|
||||
this->AddEvent(new PRS1PressureSettingEvent(PRS1_SETTING_IPAP_MAX, max_ipap, GAIN));
|
||||
break;
|
||||
case 0x19: // Tidal Volume (AVAPS)
|
||||
CHECK_VALUE(data[pos], 47); // gain 10.0
|
||||
//CHECK_VALUE(data[pos], 47); // gain 10.0
|
||||
// TODO: add a setting for this
|
||||
break;
|
||||
case 0x1e: // Backup rate (S/T and AVAPS)
|
||||
CHECK_VALUES(cpapmode, MODE_ASV, MODE_AVAPS);
|
||||
|
Loading…