Add support for PRS1 humidifier passover and error modes based on new data.
Also mark 500X180 as tested.
This commit is contained in:
parent
7ffa345ae3
commit
5e2b51c6b9
@ -222,7 +222,7 @@ enum FlexMode { FLEX_None, FLEX_CFlex, FLEX_CFlexPlus, FLEX_AFlex, FLEX_RiseTime
|
|||||||
|
|
||||||
enum BackupBreathMode { PRS1Backup_Off, PRS1Backup_Auto, PRS1Backup_Fixed };
|
enum BackupBreathMode { PRS1Backup_Off, PRS1Backup_Auto, PRS1Backup_Fixed };
|
||||||
|
|
||||||
enum HumidMode { HUMID_Fixed, HUMID_Adaptive, HUMID_HeatedTube };
|
enum HumidMode { HUMID_Fixed, HUMID_Adaptive, HUMID_HeatedTube, HUMID_Passover, HUMID_Error };
|
||||||
|
|
||||||
ChannelID PRS1_Mode = 0;
|
ChannelID PRS1_Mode = 0;
|
||||||
ChannelID PRS1_TimedBreath = 0, PRS1_HumidMode = 0, PRS1_TubeTemp = 0;
|
ChannelID PRS1_TimedBreath = 0, PRS1_HumidMode = 0, PRS1_TubeTemp = 0;
|
||||||
@ -275,6 +275,7 @@ static const PRS1TestedModel s_PRS1TestedModels[] = {
|
|||||||
{ "500X120", 0, 6, "DreamStation Auto CPAP" },
|
{ "500X120", 0, 6, "DreamStation Auto CPAP" },
|
||||||
{ "500X130", 0, 6, "DreamStation Auto CPAP" },
|
{ "500X130", 0, 6, "DreamStation Auto CPAP" },
|
||||||
{ "500X150", 0, 6, "DreamStation Auto CPAP" },
|
{ "500X150", 0, 6, "DreamStation Auto CPAP" },
|
||||||
|
{ "500X180", 0, 6, "DreamStation Auto CPAP" },
|
||||||
{ "501X120", 0, 6, "DreamStation Auto CPAP with P-Flex" },
|
{ "501X120", 0, 6, "DreamStation Auto CPAP with P-Flex" },
|
||||||
{ "500G110", 0, 6, "DreamStation Go Auto" },
|
{ "500G110", 0, 6, "DreamStation Go Auto" },
|
||||||
{ "502G150", 0, 6, "DreamStation Go Auto" },
|
{ "502G150", 0, 6, "DreamStation Go Auto" },
|
||||||
@ -6555,9 +6556,12 @@ void PRS1DataChunk::ParseTubingTypeV3(unsigned char type)
|
|||||||
|
|
||||||
void PRS1DataChunk::ParseHumidifierSettingV3(unsigned char byte1, unsigned char byte2, bool add_setting)
|
void PRS1DataChunk::ParseHumidifierSettingV3(unsigned char byte1, unsigned char byte2, bool add_setting)
|
||||||
{
|
{
|
||||||
|
bool humidifier_present = true;
|
||||||
bool humidfixed = false; // formerly called "Classic"
|
bool humidfixed = false; // formerly called "Classic"
|
||||||
bool humidadaptive = false; // formerly called "System One"
|
bool humidadaptive = false; // formerly called "System One"
|
||||||
bool tubepresent = false;
|
bool tubepresent = false;
|
||||||
|
bool passover = false;
|
||||||
|
bool error = false;
|
||||||
|
|
||||||
// Byte 1: 0x90 (no humidifier data), 0x50 (15ht, tube 4/5, humid 4), 0x54 (15ht, tube 5, humid 5) 0x4c (15ht, tube temp 3, humidifier 3)
|
// Byte 1: 0x90 (no humidifier data), 0x50 (15ht, tube 4/5, humid 4), 0x54 (15ht, tube 5, humid 5) 0x4c (15ht, tube temp 3, humidifier 3)
|
||||||
// 0x0c (15, tube 3, humid 3, fixed)
|
// 0x0c (15, tube 3, humid 3, fixed)
|
||||||
@ -6565,6 +6569,8 @@ void PRS1DataChunk::ParseHumidifierSettingV3(unsigned char byte1, unsigned char
|
|||||||
// 0b0101 0000 tube 4 and 5, humidifier 4
|
// 0b0101 0000 tube 4 and 5, humidifier 4
|
||||||
// 0b0101 0100 15ht, tube 5, humidifier 5
|
// 0b0101 0100 15ht, tube 5, humidifier 5
|
||||||
// 0b0100 1100 15ht, tube 3, humidifier 3
|
// 0b0100 1100 15ht, tube 3, humidifier 3
|
||||||
|
// 0b1011 0000 15, tube 3, humidifier 3, "Error" on humidification chart with asterisk at 4
|
||||||
|
// 0b0111 0000 15, tube 3, humidifier 3, "Passover" on humidification chart with notch at 4
|
||||||
// 842 = humidifier status
|
// 842 = humidifier status
|
||||||
// 1 84 = humidifier setting
|
// 1 84 = humidifier setting
|
||||||
// ??
|
// ??
|
||||||
@ -6574,12 +6580,13 @@ void PRS1DataChunk::ParseHumidifierSettingV3(unsigned char byte1, unsigned char
|
|||||||
case 0: humidfixed = true; break; // fixed, ignores tubetemp bits and reports tubetemp=3
|
case 0: humidfixed = true; break; // fixed, ignores tubetemp bits and reports tubetemp=3
|
||||||
case 1: humidadaptive = true; break; // adaptive, ignores tubetemp bits and reports tubetemp=3
|
case 1: humidadaptive = true; break; // adaptive, ignores tubetemp bits and reports tubetemp=3
|
||||||
case 2: tubepresent = true; break; // heated tube
|
case 2: tubepresent = true; break; // heated tube
|
||||||
case 4: break; // no humidifier, possibly a bit flag rather than integer value, reports tubetemp=3 and humidlevel=3
|
case 3: passover = true; break; // passover mode (only visible in chart)
|
||||||
|
case 4: humidifier_present = false; break; // no humidifier, reports tubetemp=3 and humidlevel=3
|
||||||
|
case 5: error = true; break; // "Error" in humidification chart, reports tubetemp=3 and humidlevel=3 in settings
|
||||||
default:
|
default:
|
||||||
UNEXPECTED_VALUE(humid, "known value");
|
UNEXPECTED_VALUE(humid, "known value");
|
||||||
break;
|
break;
|
||||||
}
|
}
|
||||||
bool humidifier_present = ((byte1 & 0x80) == 0);
|
|
||||||
int humidlevel = (byte1 >> 2) & 7;
|
int humidlevel = (byte1 >> 2) & 7;
|
||||||
|
|
||||||
// Byte 2: 0xB4 (15ht, tube 5, humid 5), 0xB0 (15ht, tube 5, humid 4), 0x90 (tube 4, humid 4), 0x6C (15ht, tube temp 3, humidifier 3)
|
// Byte 2: 0xB4 (15ht, tube 5, humid 5), 0xB0 (15ht, tube 5, humid 4), 0x90 (tube 4, humid 4), 0x6C (15ht, tube temp 3, humidifier 3)
|
||||||
@ -6602,11 +6609,16 @@ void PRS1DataChunk::ParseHumidifierSettingV3(unsigned char byte1, unsigned char
|
|||||||
}
|
}
|
||||||
}
|
}
|
||||||
|
|
||||||
|
// TODO: move this up into the switch statement above, given how many modes there now are.
|
||||||
HumidMode humidmode = HUMID_Fixed;
|
HumidMode humidmode = HUMID_Fixed;
|
||||||
if (tubepresent) {
|
if (tubepresent) {
|
||||||
humidmode = HUMID_HeatedTube;
|
humidmode = HUMID_HeatedTube;
|
||||||
} else if (humidadaptive) {
|
} else if (humidadaptive) {
|
||||||
humidmode = HUMID_Adaptive;
|
humidmode = HUMID_Adaptive;
|
||||||
|
} else if (passover) {
|
||||||
|
humidmode = HUMID_Passover;
|
||||||
|
} else if (error) {
|
||||||
|
humidmode = HUMID_Error;
|
||||||
}
|
}
|
||||||
|
|
||||||
if (add_setting) {
|
if (add_setting) {
|
||||||
@ -8813,9 +8825,11 @@ void PRS1Loader::initChannels()
|
|||||||
QObject::tr("PRS1 Humidification Mode"),
|
QObject::tr("PRS1 Humidification Mode"),
|
||||||
QObject::tr("Humid. Mode"),
|
QObject::tr("Humid. Mode"),
|
||||||
"", LOOKUP, Qt::green));
|
"", LOOKUP, Qt::green));
|
||||||
chan->addOption(0, QObject::tr("Fixed (Classic)"));
|
chan->addOption(HUMID_Fixed, QObject::tr("Fixed (Classic)"));
|
||||||
chan->addOption(1, QObject::tr("Adaptive (System One)"));
|
chan->addOption(HUMID_Adaptive, QObject::tr("Adaptive (System One)"));
|
||||||
chan->addOption(2, QObject::tr("Heated Tube"));
|
chan->addOption(HUMID_HeatedTube, QObject::tr("Heated Tube"));
|
||||||
|
chan->addOption(HUMID_Passover, QObject::tr("Passover"));
|
||||||
|
chan->addOption(HUMID_Error, QObject::tr("Error"));
|
||||||
|
|
||||||
channel.add(GRP_CPAP, chan = new Channel(PRS1_TubeTemp = 0xe10f, SETTING, MT_CPAP, SESSION,
|
channel.add(GRP_CPAP, chan = new Channel(PRS1_TubeTemp = 0xe10f, SETTING, MT_CPAP, SESSION,
|
||||||
"PRS1TubeTemp",
|
"PRS1TubeTemp",
|
||||||
|
Loading…
Reference in New Issue