Remove an unexpected data alert based on new PRS1 sample data.
In this specific instance, replace it with a warning in the debug log.
This commit is contained in:
parent
3f649fc0f5
commit
ecc6bad213
@ -653,7 +653,12 @@ void PRS1DataChunk::ParseHumidifierSetting60Series(unsigned char humid1, unsigne
|
||||
// When no_data, reports always say "System One" with humidity level 3, regardless of humidlevel and humidsystemone
|
||||
|
||||
if (humidsystemone + tubepresent + no_data == 0) CHECK_VALUE(humidclassic, true); // Always set when everything else is off
|
||||
if (no_data && humidsystemone && add_setting == false) {
|
||||
// This has been seen once on a 560P in a session that also generated a file in the error directory.
|
||||
qWarning() << this->sessionid << "Humidification error during session?";
|
||||
} else {
|
||||
if (humidsystemone + tubepresent + no_data > 1) UNEXPECTED_VALUE(humid2, "one bit set"); // Only one of these ever seems to be set at a time
|
||||
}
|
||||
if (tubepresent && tubetemp == 0) CHECK_VALUE(tubehumidlevel, 0); // When the heated tube is off, tube humidity seems to be 0
|
||||
|
||||
if (tubepresent) humidclassic = false; // Classic mode bit is evidently ignored when tube is present
|
||||
|