Improve PRS1 CPAP-Check and Auto-Trial setting import.
Also clean up PRS1 channels given the LOOKUP behavior, and clean up ImportSummary now that native PRS1 modes are supported.
This commit is contained in:
parent
3749a73fd1
commit
f49ba51cb7
@ -6734,10 +6734,8 @@ bool PRS1DataChunk::ParseSettingsF0V6(const unsigned char* data, int size)
|
||||
// The time of change is most likely in the events file. See slice 6 for ending pressure.
|
||||
//CHECK_VALUE(pressure, 0x5a);
|
||||
this->AddEvent(new PRS1PressureSettingEvent(PRS1_SETTING_PRESSURE, pressure));
|
||||
// TODO: Once OSCAR can handle more modes, we can include these settings; right now including
|
||||
// these settings makes it think this is AutoCPAP.
|
||||
//this->AddEvent(new PRS1PressureSettingEvent(PRS1_SETTING_PRESSURE_MIN, min_pressure));
|
||||
//this->AddEvent(new PRS1PressureSettingEvent(PRS1_SETTING_PRESSURE_MAX, max_pressure));
|
||||
this->AddEvent(new PRS1PressureSettingEvent(PRS1_SETTING_PRESSURE_MIN, min_pressure));
|
||||
this->AddEvent(new PRS1PressureSettingEvent(PRS1_SETTING_PRESSURE_MAX, max_pressure));
|
||||
break;
|
||||
case 0x0d: // AutoCPAP pressure setting
|
||||
CHECK_VALUE(len, 2);
|
||||
@ -6770,12 +6768,16 @@ bool PRS1DataChunk::ParseSettingsF0V6(const unsigned char* data, int size)
|
||||
this->AddEvent(new PRS1PressureSettingEvent(PRS1_SETTING_PS_MAX, imax_ps));
|
||||
break;
|
||||
case 0x10: // Auto-Trial mode
|
||||
// TODO: F0V4 considers this a separate mode from CPAP or CPAPCHECK, should F0V6 as well?
|
||||
// TODO: Check how auto-trial sessions are labeled in F0V6 reports.
|
||||
// This is not encoded as a separate mode as in F0V4, but instead as an auto-trial
|
||||
// duration on top of the CPAP or CPAP-Check mode. Reports show Auto-CPAP results,
|
||||
// but curiously report the use of C-Flex+, even though Auto-CPAP uses A-Flex.
|
||||
CHECK_VALUE(len, 3);
|
||||
CHECK_VALUES(cpapmode, PRS1_MODE_CPAP, PRS1_MODE_CPAPCHECK); // TODO: What's the difference between auto-trial and CPAP-Check?
|
||||
CHECK_VALUES(cpapmode, PRS1_MODE_CPAP, PRS1_MODE_CPAPCHECK);
|
||||
CHECK_VALUES(data[pos], 30, 5); // Auto-Trial Duration
|
||||
this->AddEvent(new PRS1ParsedSettingEvent(PRS1_SETTING_AUTO_TRIAL, data[pos]));
|
||||
// If we want C-Flex+ to be reported as A-Flex, we can set cpapmode = PRS1_MODE_AUTOTRIAL here.
|
||||
// (Note that the setting event has already been added above, which is why ImportSummary needs
|
||||
// to adjust it when it sees this setting.)
|
||||
min_pressure = data[pos+1];
|
||||
max_pressure = data[pos+2];
|
||||
this->AddEvent(new PRS1PressureSettingEvent(PRS1_SETTING_PRESSURE_MIN, min_pressure));
|
||||
@ -7517,6 +7519,7 @@ bool PRS1Import::ImportSummary()
|
||||
qint64 start = qint64(summary->timestamp) * 1000L;
|
||||
session->set_first(start);
|
||||
|
||||
// TODO: The below max pressures aren't right for the 30 cmH2O models.
|
||||
session->setPhysMax(CPAP_LeakTotal, 120);
|
||||
session->setPhysMin(CPAP_LeakTotal, 0);
|
||||
session->setPhysMax(CPAP_Pressure, 25);
|
||||
@ -7531,6 +7534,7 @@ bool PRS1Import::ImportSummary()
|
||||
bool ok;
|
||||
ok = summary->ParseSummary();
|
||||
|
||||
PRS1Mode nativemode = PRS1_MODE_UNKNOWN;
|
||||
CPAPMode cpapmode = MODE_UNKNOWN;
|
||||
for (int i=0; i < summary->m_parsedData.count(); i++) {
|
||||
PRS1ParsedEvent* e = summary->m_parsedData.at(i);
|
||||
@ -7544,7 +7548,7 @@ bool PRS1Import::ImportSummary()
|
||||
PRS1ParsedSettingEvent* s = (PRS1ParsedSettingEvent*) e;
|
||||
switch (s->m_setting) {
|
||||
case PRS1_SETTING_CPAP_MODE:
|
||||
session->settings[PRS1_Mode] = (PRS1Mode) e->m_value;
|
||||
nativemode = (PRS1Mode) e->m_value;
|
||||
cpapmode = importMode(e->m_value);
|
||||
break;
|
||||
case PRS1_SETTING_PRESSURE:
|
||||
@ -7552,11 +7556,6 @@ bool PRS1Import::ImportSummary()
|
||||
break;
|
||||
case PRS1_SETTING_PRESSURE_MIN:
|
||||
session->settings[CPAP_PressureMin] = e->value();
|
||||
if (cpapmode == MODE_CPAP) { // Auto-Trial is reported as CPAP but with a minimum and maximum pressure,
|
||||
cpapmode = MODE_APAP; // so import it as APAP, since that's what it's really doing.
|
||||
}
|
||||
// CPAP-Check no longer reports min/max pressures, since it should be treated as CPAP.
|
||||
// (It only adjusts the CPAP pressure by 1 cmH2O every 30 hours at most.)
|
||||
break;
|
||||
case PRS1_SETTING_PRESSURE_MAX:
|
||||
session->settings[CPAP_PressureMax] = e->value();
|
||||
@ -7578,9 +7577,6 @@ bool PRS1Import::ImportSummary()
|
||||
break;
|
||||
case PRS1_SETTING_IPAP_MIN:
|
||||
session->settings[CPAP_IPAPLo] = e->value();
|
||||
if (cpapmode == MODE_BILEVEL_FIXED) {
|
||||
cpapmode = MODE_BILEVEL_AUTO_VARIABLE_PS; // TODO: this isn't quite right, on ventilators it's actually fixed EPAP with variable PS
|
||||
}
|
||||
break;
|
||||
case PRS1_SETTING_IPAP_MAX:
|
||||
session->settings[CPAP_IPAPHi] = e->value();
|
||||
@ -7657,8 +7653,10 @@ bool PRS1Import::ImportSummary()
|
||||
case PRS1_SETTING_TIDAL_VOLUME:
|
||||
session->settings[CPAP_TidalVolume] = e->m_value;
|
||||
break;
|
||||
case PRS1_SETTING_AUTO_TRIAL:
|
||||
case PRS1_SETTING_AUTO_TRIAL: // new to F0V6
|
||||
session->settings[PRS1_AutoTrial] = e->m_value;
|
||||
nativemode = PRS1_MODE_AUTOTRIAL; // Note: F0V6 reports show the underlying CPAP mode rather than Auto-Trial.
|
||||
cpapmode = importMode(nativemode);
|
||||
break;
|
||||
case PRS1_SETTING_EZ_START:
|
||||
session->settings[PRS1_EZStart] = (bool) e->m_value;
|
||||
@ -7684,8 +7682,14 @@ bool PRS1Import::ImportSummary()
|
||||
if (!ok) {
|
||||
return false;
|
||||
}
|
||||
session->settings[CPAP_Mode] = cpapmode;
|
||||
|
||||
if (summary->m_parsedData.count() > 0) {
|
||||
if (nativemode == PRS1_MODE_UNKNOWN) UNEXPECTED_VALUE(nativemode, "known mode");
|
||||
if (cpapmode == MODE_UNKNOWN) UNEXPECTED_VALUE(cpapmode, "known mode");
|
||||
session->settings[PRS1_Mode] = nativemode;
|
||||
session->settings[CPAP_Mode] = cpapmode;
|
||||
}
|
||||
|
||||
if (summary->duration == 0) {
|
||||
// This does occasionally happen and merely indicates a brief session with no useful data.
|
||||
// This requires the use of really_set_last below, which otherwise rejects 0 length.
|
||||
@ -8758,13 +8762,7 @@ void PRS1Loader::initChannels()
|
||||
QObject::tr("PRS1 pressure relief setting."),
|
||||
QObject::tr("Flex Level"),
|
||||
"", LOOKUP, Qt::blue));
|
||||
|
||||
chan->addOption(0, STR_TR_Off);
|
||||
chan->addOption(1, QObject::tr("1"));
|
||||
chan->addOption(2, QObject::tr("2"));
|
||||
chan->addOption(3, QObject::tr("3"));
|
||||
chan->addOption(4, QObject::tr("4"));
|
||||
chan->addOption(5, QObject::tr("5"));
|
||||
|
||||
channel.add(GRP_CPAP, chan = new Channel(PRS1_FlexLock = 0xe111, SETTING, MT_CPAP, SESSION,
|
||||
"PRS1FlexLock",
|
||||
@ -8780,7 +8778,7 @@ void PRS1Loader::initChannels()
|
||||
QObject::tr("Rise Time"),
|
||||
QObject::tr("Amount of time it takes to transition from EPAP to IPAP, the higher the number the slower the transition"),
|
||||
QObject::tr("Rise Time"),
|
||||
"", DEFAULT, Qt::blue));
|
||||
"", LOOKUP, Qt::blue));
|
||||
|
||||
channel.add(GRP_CPAP, chan = new Channel(PRS1_RiseTimeLock = 0xe11a, SETTING, MT_CPAP, SESSION,
|
||||
"PRS1RiseTimeLock",
|
||||
@ -8817,11 +8815,6 @@ void PRS1Loader::initChannels()
|
||||
QObject::tr("Tube Temp."),
|
||||
"", LOOKUP, Qt::red));
|
||||
chan->addOption(0, STR_TR_Off);
|
||||
chan->addOption(1, QObject::tr("1"));
|
||||
chan->addOption(2, QObject::tr("2"));
|
||||
chan->addOption(3, QObject::tr("3"));
|
||||
chan->addOption(4, QObject::tr("4"));
|
||||
chan->addOption(5, QObject::tr("5"));
|
||||
|
||||
channel.add(GRP_CPAP, chan = new Channel(PRS1_HumidLevel = 0xe102, SETTING, MT_CPAP, SESSION,
|
||||
"PRS1HumidLevel",
|
||||
@ -8830,11 +8823,6 @@ void PRS1Loader::initChannels()
|
||||
QObject::tr("Humid. Lvl"),
|
||||
"", LOOKUP, Qt::blue));
|
||||
chan->addOption(0, STR_TR_Off);
|
||||
chan->addOption(1, QObject::tr("1"));
|
||||
chan->addOption(2, QObject::tr("2"));
|
||||
chan->addOption(3, QObject::tr("3"));
|
||||
chan->addOption(4, QObject::tr("4"));
|
||||
chan->addOption(5, QObject::tr("5"));
|
||||
|
||||
channel.add(GRP_CPAP, chan = new Channel(PRS1_MaskResistSet = 0xe104, SETTING, MT_CPAP, SESSION,
|
||||
"MaskResistSet",
|
||||
@ -8843,11 +8831,6 @@ void PRS1Loader::initChannels()
|
||||
QObject::tr("Mask Resist."),
|
||||
"", LOOKUP, Qt::green));
|
||||
chan->addOption(0, STR_TR_Off);
|
||||
chan->addOption(1, QObject::tr("1"));
|
||||
chan->addOption(2, QObject::tr("2"));
|
||||
chan->addOption(3, QObject::tr("3"));
|
||||
chan->addOption(4, QObject::tr("4"));
|
||||
chan->addOption(5, QObject::tr("5"));
|
||||
|
||||
channel.add(GRP_CPAP, chan = new Channel(PRS1_HoseDiam = 0xe107, SETTING, MT_CPAP, SESSION,
|
||||
"PRS1HoseDiam",
|
||||
@ -8928,7 +8911,7 @@ void PRS1Loader::initChannels()
|
||||
QObject::tr("The kind of backup breath rate in use: none (off), automatic, or fixed"),
|
||||
QObject::tr("Breath Rate"),
|
||||
"", LOOKUP, Qt::black));
|
||||
chan->addOption(0, QObject::tr("Off"));
|
||||
chan->addOption(0, STR_TR_Off);
|
||||
chan->addOption(1, QObject::tr("Auto"));
|
||||
chan->addOption(2, QObject::tr("Fixed"));
|
||||
|
||||
@ -8937,7 +8920,7 @@ void PRS1Loader::initChannels()
|
||||
QObject::tr("Fixed Backup Breath BPM"),
|
||||
QObject::tr("Minimum breaths per minute (BPM) below which a timed breath will be initiated"),
|
||||
QObject::tr("Breath BPM"),
|
||||
STR_UNIT_BreathsPerMinute, DEFAULT, Qt::black));
|
||||
STR_UNIT_BreathsPerMinute, LOOKUP, Qt::black));
|
||||
|
||||
channel.add(GRP_CPAP, chan = new Channel(PRS1_BackupBreathTi = 0xe116, SETTING, MT_CPAP, SESSION,
|
||||
"PRS1BackupBreathTi",
|
||||
@ -8948,11 +8931,10 @@ void PRS1Loader::initChannels()
|
||||
|
||||
channel.add(GRP_CPAP, chan = new Channel(PRS1_AutoTrial = 0xe117, SETTING, MT_CPAP, SESSION,
|
||||
"PRS1AutoTrial",
|
||||
QObject::tr("Auto-Trial"),
|
||||
QObject::tr("The number of days left in the Auto-CPAP trial period, after which the machine will revert to CPAP only"),
|
||||
QObject::tr("Auto-Trial"),
|
||||
QObject::tr("Auto-Trial Duration"),
|
||||
QObject::tr("The number of days in the Auto-CPAP trial period, after which the machine will revert to CPAP"),
|
||||
QObject::tr("Auto-Trial Dur."),
|
||||
"", LOOKUP, Qt::black));
|
||||
chan->addOption(0, STR_TR_Off);
|
||||
|
||||
channel.add(GRP_CPAP, chan = new Channel(PRS1_EZStart = 0xe118, SETTING, MT_CPAP, SESSION,
|
||||
"PRS1EZStart",
|
||||
|
Loading…