/* SleepLib CMS50X Loader Implementation
|
|
*
|
|
* Copyright (c) 2019-2024 The OSCAR Team
|
|
* Copyright (c) 2011-2018 Mark Watkins
|
|
*
|
|
* This file is subject to the terms and conditions of the GNU General Public
|
|
* License. See the file COPYING in the main directory of the source code
|
|
* for more details. */
|
|
|
|
//********************************************************************************************
|
|
// Please only INCREMENT the cms50f37_data_version in cms50f37_loader.h when making changes
|
|
// that change loader behaviour or modify channels in a manner that fixes old data imports.
|
|
// Note that changing the data version will require a reimport of existing data for which OSCAR
|
|
// does not keep a backup - so it should be avoided if possible.
|
|
// i.e. there is no need to change the version when adding support for new devices
|
|
//********************************************************************************************
|
|
|
|
// #include <QProgressBar>
|
|
#include <QApplication>
|
|
#include <QDir>
|
|
#include <QString>
|
|
#include <QDataStream>
|
|
#include <QDateTime>
|
|
#include <QFile>
|
|
#include <QDebug>
|
|
#include <QList>
|
|
#include <QMessageBox>
|
|
#include <QLabel>
|
|
#include <QVBoxLayout>
|
|
#include <QPushButton>
|
|
|
|
// The qt5.15 obsolescence of hex requires this change.
|
|
// this solution to QT's obsolescence is only used in debug statements
|
|
#if QT_VERSION >= QT_VERSION_CHECK(5,15,0)
|
|
#define QTHEX Qt::hex
|
|
#define QTDEC Qt::dec
|
|
#else
|
|
#define QTHEX hex
|
|
#define QTDEC dec
|
|
#endif
|
|
|
|
using namespace std;
|
|
|
|
#include "cms50f37_loader.h"
|
|
#include "SleepLib/machine.h"
|
|
#include "SleepLib/session.h"
|
|
|
|
CMS50F37Loader::CMS50F37Loader()
|
|
{
|
|
m_type = MT_OXIMETER;
|
|
m_abort = false;
|
|
m_streaming = false;
|
|
m_importing = false;
|
|
started_reading = false;
|
|
cms50dplus = false;
|
|
|
|
imp_callbacks = 0;
|
|
|
|
m_vendorID = 0x10c4;
|
|
m_productID = 0xea60;
|
|
|
|
oxirec = nullptr;
|
|
|
|
startTimer.setParent(this);
|
|
resetTimer.setParent(this);
|
|
|
|
duration_divisor = 2;
|
|
|
|
model = QString();
|
|
vendor = QString();
|
|
|
|
}
|
|
|
|
CMS50F37Loader::~CMS50F37Loader()
|
|
{
|
|
}
|
|
|
|
bool CMS50F37Loader::openDevice()
|
|
{
|
|
if (port.isEmpty()) {
|
|
bool b = scanDevice("",m_vendorID, m_productID);
|
|
if (!b) {
|
|
b = scanDevice("rfcomm", 0, 0); // Linux
|
|
}
|
|
if (!b) {
|
|
qWarning() << "cms50f37 - No oximeter found";
|
|
return false;
|
|
}
|
|
}
|
|
serial.setPortName(port);
|
|
if (!serial.open(QSerialPort::ReadWrite)) {
|
|
qDebug() << "cms50f37 - Failed to open oximeter";
|
|
return false;
|
|
}
|
|
|
|
// forward this stuff
|
|
|
|
// Set up serial port attributes
|
|
serial.setBaudRate(QSerialPort::Baud115200);
|
|
serial.setParity(QSerialPort::NoParity);
|
|
serial.setStopBits(QSerialPort::OneStop);
|
|
serial.setDataBits(QSerialPort::Data8);
|
|
serial.setFlowControl(QSerialPort::NoFlowControl);
|
|
|
|
m_streaming = true;
|
|
m_abort = false;
|
|
m_importing = false;
|
|
|
|
// connect relevant signals
|
|
connect(&serial,SIGNAL(readyRead()), this, SLOT(dataAvailable()));
|
|
|
|
started_import = false;
|
|
started_reading = false;
|
|
finished_import = false;
|
|
|
|
resetDevice();
|
|
return true;
|
|
}
|
|
|
|
bool CMS50F37Loader::Detect(const QString &path)
|
|
{
|
|
if (p_profile->oxi->oximeterType() == 0) {
|
|
return true;
|
|
}
|
|
Q_UNUSED(path);
|
|
return false;
|
|
}
|
|
|
|
int CMS50F37Loader::Open(const QString & path)
|
|
{
|
|
// Only one active Oximeter module at a time, set in preferences
|
|
|
|
m_itemCnt = 0;
|
|
m_itemTotal = 0;
|
|
|
|
m_abort = false;
|
|
m_importing = false;
|
|
|
|
started_import = false;
|
|
started_reading = false;
|
|
finished_import = false;
|
|
setStatus(NEUTRAL);
|
|
|
|
imp_callbacks = 0;
|
|
cb_reset = 0;
|
|
|
|
m_time.start();
|
|
|
|
if (oxirec) {
|
|
trashRecords();
|
|
}
|
|
|
|
// Cheating using path for two serial oximetry modes
|
|
|
|
if (path.compare("import") == 0) {
|
|
serial.clear();
|
|
|
|
sequence = 0;
|
|
buffer.clear();
|
|
|
|
// nextCommand();
|
|
setStatus(IMPORTING);
|
|
|
|
return 1;
|
|
} else if (path.compare("live") == 0) {
|
|
return 0;
|
|
}
|
|
QString ext = path.section(".", -1); // find the LAST '.'
|
|
if ((ext.compare("spo2", Qt::CaseInsensitive)==0) || (ext.compare("spo", Qt::CaseInsensitive)==0) || (ext.compare("spor", Qt::CaseInsensitive)==0)) {
|
|
// try to read and process SpoR file..
|
|
return readSpoRFile(path) ? 1 : 0;
|
|
}
|
|
|
|
return 0;
|
|
}
|
|
|
|
unsigned char cms50_sequence[] = { 0xa7, 0xa2, 0xa0, 0xb0, 0xac, 0xb3, 0xad, 0xa3, 0xab, 0xa4, 0xa5, 0xaf, 0xa7, 0xa2, 0xa6 };
|
|
|
|
const int TIMEOUT = 2000;
|
|
|
|
|
|
//const quint8 COMMAND_GET_VERSION = 0xA0; // not sure of this one
|
|
const quint8 COMMAND_CMS50_HELLO2 = 0xA2; // stop live data stream
|
|
const quint8 COMMAND_GET_SESSION_COUNT = 0xA3;
|
|
const quint8 COMMAND_GET_SESSION_DURATION = 0xA4;
|
|
const quint8 COMMAND_GET_SESSION_TIME = 0xA5;
|
|
const quint8 COMMAND_GET_SESSION_DATA = 0xA6;
|
|
const quint8 COMMAND_CMS50_HELLO1 = 0xA7; // stop stored data stream
|
|
const quint8 COMMAND_GET_OXIMETER_MODEL = 0xA8;
|
|
const quint8 COMMAND_GET_OXIMETER_VENDOR = 0xA9;
|
|
const quint8 COMMAND_GET_OXIMETER_DEVICEID = 0xAA;
|
|
const quint8 COMMAND_GET_USER_INFO = 0xAB;
|
|
const quint8 COMMAND_GET_USER_COUNT = 0xAD; // for future check and use
|
|
const quint8 COMMAND_SESSION_ERASE = 0xAE;
|
|
const quint8 COMMAND_GET_OXIMETER_INFO = 0xB0;
|
|
|
|
|
|
int cms50_seqlength = sizeof(cms50_sequence);
|
|
|
|
|
|
QString CMS50F37Loader::getUser()
|
|
{
|
|
if (!userName.isEmpty()) return userName;
|
|
|
|
sendCommand(COMMAND_GET_USER_INFO);
|
|
|
|
QElapsedTimer time;
|
|
time.start();
|
|
do {
|
|
QApplication::processEvents();
|
|
} while (userName.isEmpty() && (time.elapsed() < TIMEOUT));
|
|
|
|
qDebug() << "cms50f37 - User " << userIdx << " is " << userName;
|
|
return userName;
|
|
}
|
|
|
|
QString CMS50F37Loader::getVendor()
|
|
{
|
|
if (!vendor.isEmpty()) return vendor;
|
|
|
|
sendCommand(COMMAND_GET_OXIMETER_VENDOR);
|
|
|
|
QElapsedTimer time;
|
|
time.start();
|
|
do {
|
|
QApplication::processEvents();
|
|
} while (vendor.isEmpty() && (time.elapsed() < TIMEOUT));
|
|
|
|
qDebug() << "cms50f37 - Vendor is " << vendor;
|
|
return vendor;
|
|
}
|
|
|
|
QString CMS50F37Loader::getModel()
|
|
{
|
|
if (!model.isEmpty()) return model;
|
|
|
|
modelsegments = 0;
|
|
sendCommand(COMMAND_GET_OXIMETER_MODEL);
|
|
|
|
QElapsedTimer time;
|
|
time.start();
|
|
do {
|
|
QApplication::processEvents();
|
|
} while ((modelsegments < 2) && (time.elapsed() < TIMEOUT));
|
|
|
|
// Give a little more time for the second one..
|
|
QThread::msleep(100);
|
|
QApplication::processEvents();
|
|
|
|
if (model.startsWith("CMS50I") || model.startsWith("CMS50H")) {
|
|
duration_divisor = 4;
|
|
} else {
|
|
duration_divisor = 2;
|
|
}
|
|
|
|
qDebug() << "cms50f37 - Model is " << model;
|
|
return model;
|
|
}
|
|
|
|
QString CMS50F37Loader::getDeviceString()
|
|
{
|
|
QString VendDev = QString("%1 %2").arg(getVendor()).arg(getModel());
|
|
qDebug() << "cms50f37 - USB Device String is " << VendDev;
|
|
return VendDev;
|
|
}
|
|
|
|
QString CMS50F37Loader::getDeviceID()
|
|
{
|
|
if (!devid.isEmpty()) return devid;
|
|
|
|
sendCommand(COMMAND_GET_OXIMETER_DEVICEID);
|
|
|
|
QElapsedTimer time;
|
|
time.start();
|
|
do {
|
|
QApplication::processEvents();
|
|
} while (devid.isEmpty() && (time.elapsed() < TIMEOUT));
|
|
|
|
qDebug() << "cms50f37 - Device Id is " << devid;
|
|
return devid;
|
|
}
|
|
|
|
int CMS50F37Loader::getUserCount() // for future use, check, then add select user if > 1
|
|
{
|
|
userCount = -1;
|
|
sendCommand(COMMAND_GET_USER_COUNT);
|
|
|
|
QElapsedTimer time;
|
|
time.start();
|
|
do {
|
|
QApplication::processEvents();
|
|
} while ((userCount < 0) && (time.elapsed() < TIMEOUT));
|
|
|
|
qDebug() << "cms50f37 - User count is " << userCount;
|
|
return userCount;
|
|
}
|
|
|
|
int CMS50F37Loader::getSessionCount()
|
|
{
|
|
session_count = -1;
|
|
sendCommand(COMMAND_GET_SESSION_COUNT);
|
|
|
|
QElapsedTimer time;
|
|
time.start();
|
|
do {
|
|
QApplication::processEvents();
|
|
} while ((session_count < 0) && (time.elapsed() < TIMEOUT));
|
|
|
|
qDebug() << "cms50f37 - Session count is " << session_count;
|
|
return session_count;
|
|
}
|
|
|
|
int CMS50F37Loader::getOximeterInfo()
|
|
{
|
|
device_info = -1;
|
|
sendCommand(COMMAND_GET_OXIMETER_INFO);
|
|
QElapsedTimer time;
|
|
time.start();
|
|
do {
|
|
QApplication::processEvents();
|
|
} while ((device_info < 0) && (time.elapsed() < TIMEOUT));
|
|
|
|
qDebug() << "cms50f37 - Device Info is " << device_info;
|
|
return device_info;
|
|
}
|
|
|
|
int CMS50F37Loader::getDuration(int session)
|
|
{
|
|
getOximeterInfo();
|
|
|
|
duration = -1;
|
|
sendCommand(COMMAND_GET_SESSION_DURATION, session);
|
|
|
|
QElapsedTimer time;
|
|
time.start();
|
|
do {
|
|
QApplication::processEvents();
|
|
} while ((duration < 0) && (time.elapsed() < TIMEOUT));
|
|
|
|
qDebug() << "cms50f37 - Session duration is " << duration << "Divided by " << duration_divisor;
|
|
return duration / duration_divisor;
|
|
}
|
|
|
|
|
|
QDateTime CMS50F37Loader::getDateTime(int session)
|
|
{
|
|
QDateTime datetime;
|
|
imp_date = QDate();
|
|
imp_time = QTime();
|
|
sendCommand(COMMAND_GET_SESSION_TIME, session);
|
|
QElapsedTimer time;
|
|
time.start();
|
|
do {
|
|
QApplication::processEvents();
|
|
} while ((imp_date.isNull() || imp_time.isNull()) && (time.elapsed() < TIMEOUT));
|
|
|
|
if (imp_date.isNull() && imp_time.isNull())
|
|
datetime = QDateTime(QDate::currentDate(), QTime::currentTime());
|
|
else if ( imp_date.isNull() )
|
|
datetime = QDateTime(QDate::currentDate(), imp_time);
|
|
else if ( imp_time.isNull() )
|
|
datetime = QDateTime(imp_date, QTime::currentTime());
|
|
else
|
|
datetime = QDateTime(imp_date, imp_time);
|
|
|
|
qDebug() << "cms50f37 - Oximeter DateTime is " << datetime.toString("yyyy-MMM-dd HH:mm:ssap");
|
|
return datetime;
|
|
}
|
|
|
|
|
|
void CMS50F37Loader::processBytes(QByteArray bytes)
|
|
{
|
|
static quint8 resimport = 0;
|
|
int data;
|
|
|
|
QString tmpstr;
|
|
|
|
int lengths[32] = { 0, 0, 9, 9, 9, 9, 4, 8, 8, 6, 4, 4, 2, 0, 3, 8, 3, 9, 8, 9, 9, 0, 0, 0, 0, 0, 0, 0, 0, 0, 0, 0 };
|
|
|
|
buffer.append(bytes);
|
|
int size = buffer.size();
|
|
|
|
int idx = 0;
|
|
int len;
|
|
|
|
int year, month, day;
|
|
|
|
quint8 pulse;
|
|
quint8 spo2;
|
|
quint16 pi; // perfusion index
|
|
|
|
do {
|
|
quint8 res = buffer.at(idx);
|
|
|
|
len = lengths[res & 0x1f];
|
|
|
|
if ((idx+len) > size)
|
|
break;
|
|
|
|
if (started_reading && (res != resimport)) {
|
|
len = 0;
|
|
}
|
|
|
|
if (len == 0) {
|
|
// lost sync
|
|
if (started_reading) {
|
|
while (idx < size) {
|
|
res = buffer.at(idx++);
|
|
if (res == resimport) break;
|
|
}
|
|
// add a dummy to make up for it.
|
|
qDebug() << "cms50f37 - pB: lost sync, padding...";
|
|
oxirec->append(OxiRecord(0,0,0));
|
|
}
|
|
continue;
|
|
}
|
|
|
|
// always copy the high bits into the following bytes
|
|
for (int i = 2, msb = buffer.at(idx+1); i < len; i++, msb>>= 1) {
|
|
buffer[idx+i] = (buffer[idx+i] & 0x7f) | (msb & 0x01 ? 0x80 : 0);
|
|
}
|
|
|
|
if (!started_reading) switch(res) {
|
|
case 0x02: // Model name string (there are two in sequnce.. second might be the next chunk!)
|
|
data = buffer.at(idx+2);
|
|
if (data == 0) {
|
|
model = QString(buffer.mid(idx+3, 6));
|
|
modelsegments++;
|
|
} else {
|
|
QString extra = QString(buffer.mid(idx+3, 6));
|
|
model += extra.trimmed();
|
|
modelsegments++;
|
|
qDebug() << "cms50f37 - pB: Model:" << model;
|
|
}
|
|
break;
|
|
case 0x03: // Vendor string
|
|
vendor = QString(buffer.mid(idx+2, 7));
|
|
qDebug() << "cms50f37 - pB: Vendor:" << vendor;
|
|
break;
|
|
|
|
case 0x04: // Device Identifiers
|
|
devid = QString(buffer.mid(idx+2, 7));
|
|
qDebug() << "cms50f37 - pB: Device ID:" << devid;
|
|
break;
|
|
|
|
// COMMAND_GET_USER_INFO
|
|
case 0x05: // 5,80,80,f5,f3,e5,f2,80,80
|
|
userIdx = buffer.at(idx+2); // for future use
|
|
userName = QString(buffer.mid(idx+3).trimmed());
|
|
qDebug() << "cms50f37 - pB: 0x05:" << userName;
|
|
break;
|
|
|
|
// Command GET_VERSION
|
|
case 0x6: // 6,80,80,87
|
|
protocolVersion = buffer.at(idx+3) | (buffer.at(idx+2)<<8);
|
|
break;
|
|
|
|
// COMMAND_GET_SESSION_TIME --- the date part
|
|
case 0x07: // 7,80,80,80,94,8e,88,92
|
|
|
|
year = QString::asprintf("%02i%02i",buffer.at(idx+4), buffer.at(idx+5)).toInt();
|
|
month = QString::asprintf("%02i", buffer.at(idx+6)).toInt();
|
|
day = QString::asprintf("%02i", buffer.at(idx+7)).toInt();
|
|
|
|
if ( year == 0 ) {
|
|
imp_date = QDate::currentDate();
|
|
cms50dplus = true;
|
|
}
|
|
else
|
|
imp_date = QDate(year,month,day);
|
|
qDebug() << "cms50f37 - cms50D+ detected: " << (cms50dplus ? "yes" : "no");
|
|
qDebug() << "cms50f37 - pB: ymd " << year << month << day << " impDate: " << imp_date;
|
|
break;
|
|
|
|
// COMMAND_GET_SESSION_DURATION
|
|
case 0x08: // 8,80,80,80,a4,81,80,80 // 00, 00, 24, 01, 00, 00
|
|
|
|
duration = buffer.at(idx+4);
|
|
duration |= (buffer.at(idx+5) << 8);
|
|
duration |= (buffer.at(idx+6) << 16);
|
|
duration |= (buffer.at(idx+7) << 24);
|
|
break;
|
|
|
|
// COMMAND_GET_SESSION_COUNT
|
|
case 0x0a: // a,80,80,81
|
|
userIdx = buffer.at(idx+2);
|
|
session_count = buffer.at(idx+3);
|
|
break;
|
|
|
|
case 0x0b:
|
|
timectr++;
|
|
break;
|
|
|
|
// COMMAND_CMS50_HELLO1 && COMMAND_CMS50_HELLO2
|
|
case 0xc: // a7 & a2 // responds with: c,80
|
|
//data = buffer.at(idx+1);
|
|
break;
|
|
|
|
// Query Perfusion Index available (0 is yes, 1 is no)
|
|
case 0x0e: // e,80,81
|
|
//data = buffer.at(idx+2);
|
|
break;
|
|
|
|
// Get User Count
|
|
case 0x10: // 10,80,81
|
|
userCount = buffer.at(idx+2);
|
|
break;
|
|
|
|
// COMMAND_GET_OXIMETER_INFO
|
|
case 0x11: // 11,80,81,81,80,80,80,80,80
|
|
device_info = buffer.at(idx+3);
|
|
break;
|
|
|
|
// COMMAND_GET_SESSION_TIME
|
|
case 0x12: // 12,80,80,80,82,a6,92,80
|
|
tmpstr = QString::asprintf("%02i:%02i:%02i",buffer.at(idx+4), buffer.at(idx+5), buffer.at(idx+6));
|
|
imp_time = QTime::fromString(tmpstr, "HH:mm:ss");
|
|
qDebug() << "cms50f37 - pB: tmpStr:" << tmpstr << " impTime: " << imp_time;
|
|
|
|
break;
|
|
case 0x13: // 13,80,a0,a0,a0,a0,a0,a0,a0
|
|
break;
|
|
case 0x14: //14,80,80,80,80,80,80,80,80
|
|
break;
|
|
|
|
case 0x09: // cms50i data sequence
|
|
case 0x0f: // f,80,de,c2,de,c2,de,c2 cms50F data...
|
|
if (!started_import) {
|
|
qDebug() << "cms50f37 - pB: Starting import";
|
|
started_import = true;
|
|
started_reading = true;
|
|
finished_import = false;
|
|
|
|
m_importing = true;
|
|
|
|
m_itemCnt=0;
|
|
m_itemTotal=duration;
|
|
|
|
have_perfindex = (res == 0x9);
|
|
|
|
oxirec = new QVector<OxiRecord>;
|
|
oxirec->reserve(30000);
|
|
|
|
oxisessions[m_startTime] = oxirec;
|
|
|
|
cb_reset = 1;
|
|
|
|
resetTimer.singleShot(2000,this,SLOT(resetImportTimeout()));
|
|
resimport = res;
|
|
}
|
|
|
|
break;
|
|
default:
|
|
qDebug() << "cms50f37 - pB: unknown cms50F result?" << QTHEX << (int)res;
|
|
break;
|
|
}
|
|
|
|
if (res == 0x09) {
|
|
// 9,80,e1,c4,ce,82 // cms50i data
|
|
|
|
pi = buffer.at(idx+4) | (buffer.at(idx+5) << 8);
|
|
pulse = buffer.at(idx+3);
|
|
spo2 = buffer.at(idx+2);
|
|
// qDebug() << "cms50f37 - pB: Pulse=" << pulse << "SPO2=" << spo2 << "PI=" << pi;
|
|
|
|
oxirec->append(((spo2 == 0) || (pulse == 0)) ? OxiRecord(0,0,0) : OxiRecord(pulse, spo2, pi));
|
|
} else if (res == 0x0f) {
|
|
// f,80,de,c2,de,c2,de,c2 cms50F data...
|
|
|
|
pulse = buffer.at(idx+3);
|
|
spo2 = buffer.at(idx+2);
|
|
// qDebug() << "cms50f37 - pB: Pulse=" << pulse << "SPO2=" << spo2;
|
|
oxirec->append((pulse == 0xff) ? OxiRecord(0,0) : OxiRecord(pulse, spo2));
|
|
|
|
pulse = buffer.at(idx+5);
|
|
spo2 = buffer.at(idx+4);
|
|
// qDebug() << "cms50f37 - pB: Pulse=" << pulse << "SPO2=" << spo2;
|
|
oxirec->append((pulse == 0xff) ? OxiRecord(0,0) : OxiRecord(pulse, spo2));
|
|
|
|
pulse = buffer.at(idx+7);
|
|
spo2 = buffer.at(idx+6);
|
|
// qDebug() << "cms50f37 - pB: Pulse=" << pulse << "SPO2=" << spo2;
|
|
oxirec->append((pulse == 0xff) ? OxiRecord(0,0) : OxiRecord(pulse, spo2));
|
|
}
|
|
|
|
QStringList str;
|
|
for (int i=0; i < len; ++i) {
|
|
str.append(QString::number((unsigned char)buffer.at(idx + i),16));
|
|
}
|
|
|
|
if (!started_import) {
|
|
// startTimer.singleShot(2000, this, SLOT(requestData()));
|
|
importCount = 0;
|
|
qDebug() << "cms50f37 - pB: Read:" << len << size << str.join(",");
|
|
} else {
|
|
importCount++;
|
|
// qDebug() << "cms50f37 - pB: Import:" << len << size << str.join(",");
|
|
}
|
|
|
|
idx += len;
|
|
} while (idx < size);
|
|
|
|
|
|
if (!started_import) {
|
|
imp_callbacks = 0;
|
|
} else {
|
|
emit updateProgress(oxirec->size(), duration);
|
|
|
|
imp_callbacks++;
|
|
}
|
|
|
|
buffer = buffer.mid(idx);
|
|
}
|
|
|
|
|
|
//int CMS50F37Loader::doLiveMode()
|
|
//{
|
|
// if (oxirec == nullptr) { // warn
|
|
//}
|
|
//
|
|
// int available = buffer.size();
|
|
// int idx = 0;
|
|
//
|
|
// QByteArray plethy;
|
|
// while (idx < available-5) {
|
|
// if (((unsigned char)buffer.at(idx) & 0x80) != 0x80) {
|
|
// idx++;
|
|
// continue;
|
|
// }
|
|
// int pwave=(unsigned char)buffer.at(idx + 1);
|
|
// int pbeat=(unsigned char)buffer.at(idx + 2);
|
|
// int pulse=((unsigned char)buffer.at(idx + 3) & 0x7f) | ((pbeat & 0x40) << 1);
|
|
// int spo2=(unsigned char)buffer.at(idx + 4) & 0x7f;
|
|
//
|
|
// oxirec->append(OxiRecord(pulse, spo2));
|
|
// plethy.append(pwave);
|
|
//
|
|
// idx += 5;
|
|
// }
|
|
// emit updatePlethy(plethy);
|
|
//
|
|
// return idx;
|
|
//}
|
|
|
|
|
|
void CMS50F37Loader::sendCommand(quint8 c)
|
|
{
|
|
quint8 cmd[] = { 0x7d, 0x81, 0x80, 0x80, 0x80, 0x80, 0x80, 0x80, 0x80 };
|
|
cmd[2] |= (c & 0x7f);
|
|
|
|
QString out;
|
|
for (int i=0;i < 9;i++)
|
|
out += QString::asprintf("%02X ",cmd[i]);
|
|
qDebug() << "cms50f37 - Write:" << out;
|
|
|
|
if (serial.write((char *)cmd, 9) == -1) {
|
|
qDebug() << "cms50f37 - Couldn't write data bytes to CMS50F";
|
|
}
|
|
}
|
|
|
|
void CMS50F37Loader::sendCommand(quint8 c, quint8 c2)
|
|
{
|
|
quint8 cmd[] = { 0x7d, 0x81, 0x80, 0x80, 0x80, 0x80, 0x80, 0x80, 0x80 };
|
|
cmd[2] |= (c & 0x7f);
|
|
cmd[4] |= (c2 & 0x7f);
|
|
|
|
QString out;
|
|
for (int i=0; i < 9; ++i)
|
|
out += QString::asprintf("%02X ",cmd[i]);
|
|
qDebug() << "cms50f37 - Write:" << out;
|
|
|
|
if (serial.write((char *)cmd, 9) == -1) {
|
|
qDebug() << "cms50f37 - Couldn't write data bytes to CMS50F";
|
|
}
|
|
}
|
|
|
|
void CMS50F37Loader::eraseSession(int user, int session)
|
|
{
|
|
quint8 cmd[] = { 0x7d, 0x81, COMMAND_SESSION_ERASE, 0x80, 0x80, 0x80, 0x80, 0x80, 0x80 };
|
|
|
|
cmd[3] = (user & 0x7f) | 0x80;
|
|
cmd[4] = (session & 0x7f) | 0x80;
|
|
|
|
QString out;
|
|
for (int i=0; i < 9; ++i)
|
|
out += QString::asprintf("%02X ",cmd[i]);
|
|
qDebug() << "cms50f37 - Erase Session: Write:" << out;
|
|
|
|
if (serial.write((char *)cmd, 9) == -1) {
|
|
qDebug() << "cms50f37 - Couldn't write Erase session bytes to CMS50F";
|
|
}
|
|
|
|
int z = timectr;
|
|
QElapsedTimer time;
|
|
time.start();
|
|
do {
|
|
QApplication::processEvents();
|
|
} while ((timectr == z) && (time.elapsed() < TIMEOUT));
|
|
}
|
|
|
|
|
|
void CMS50F37Loader::setDeviceID(const QString & newid)
|
|
{
|
|
QString str = newid;
|
|
str.truncate(7);
|
|
if (str.length() < 7) {
|
|
str = QString(" ").repeated(7-str.length()) + str;
|
|
}
|
|
quint8 cmd[] = { 0x04, 0x80, 0x80, 0x80, 0x80, 0x80, 0x80, 0x80, 0x80 };
|
|
|
|
quint8 msb = 0;
|
|
|
|
QByteArray ba = str.toLocal8Bit();
|
|
for (int i=6; i >= 0; i--) {
|
|
msb <<= 1;
|
|
msb |= (ba.at(i) >> 7) & 1;
|
|
cmd[i+2] = ba.at(i) | 0x80;
|
|
}
|
|
|
|
cmd[1] = msb | 0x80;
|
|
|
|
QString out;
|
|
for (int i=0; i < 9; ++i)
|
|
out += QString::asprintf("%02X ",cmd[i]);
|
|
qDebug() << "cms50f37 - setDeviceID: Write:" << out;
|
|
|
|
if (serial.write((char *)cmd, 9) == -1) {
|
|
qDebug() << "cms50f37 - Couldn't write DeviceID data bytes to CMS50F";
|
|
}
|
|
|
|
// Supposed to return 0x04 command, so reset devid..
|
|
devid = QString();
|
|
|
|
QElapsedTimer time;
|
|
time.start();
|
|
do {
|
|
QApplication::processEvents();
|
|
} while (devid.isEmpty() && (time.elapsed() < TIMEOUT));
|
|
}
|
|
|
|
void CMS50F37Loader::syncClock()
|
|
{
|
|
QDate date = QDate::currentDate();
|
|
int year = date.year();
|
|
quint8 yh = (year / 100) | 0x80;
|
|
quint8 yl = (year % 100) | 0x80;
|
|
quint8 mon = date.month() | 0x80;
|
|
quint8 day = date.day() | 0x80;
|
|
quint8 wd = (date.dayOfWeek() % 7) | 0x80;
|
|
|
|
quint8 datecmd[] = { 0x7d, 0x81, 0xb2, yh, yl, mon, day, wd, 0x80 };
|
|
|
|
timectr = 0;
|
|
if (serial.write((char *)datecmd, 9) == -1) {
|
|
qDebug() << "cms50f37 - Couldn't write date bytes to CMS50F";
|
|
}
|
|
|
|
QElapsedTimer time;
|
|
time.start();
|
|
do {
|
|
QApplication::processEvents();
|
|
} while ((timectr == 0) && (time.elapsed() < TIMEOUT));
|
|
|
|
|
|
QThread::msleep(100);
|
|
QApplication::processEvents();
|
|
|
|
QTime ctime = QTime::currentTime();
|
|
quint8 h = ctime.hour() | 0x80;
|
|
quint8 m = ctime.minute() | 0x80;
|
|
quint8 s = ctime.second() | 0x80;
|
|
|
|
quint8 timecmd[] = { 0x7d, 0x81, 0xb1, h, m, s, 0x80, 0x80, 0x80 };
|
|
|
|
timectr = 0;
|
|
if (serial.write((char *)timecmd, 9) == -1) {
|
|
qDebug() << "cms50f37 - Couldn't write time bytes to CMS50F";
|
|
}
|
|
|
|
time.start();
|
|
do {
|
|
QApplication::processEvents();
|
|
} while ((timectr == 0) && (time.elapsed() < TIMEOUT));
|
|
}
|
|
|
|
|
|
void CMS50F37Loader::nextCommand()
|
|
{
|
|
qDebug() << "nextCommand sequence: " << sequence;
|
|
if (++sequence < cms50_seqlength) {
|
|
// Send the next command packet in sequence
|
|
sendCommand(cms50_sequence[sequence]);
|
|
} else {
|
|
qDebug() << "cms50f37 - Run out of startup tasks to do and import failed!";
|
|
}
|
|
}
|
|
|
|
void CMS50F37Loader::getSessionData(int session)
|
|
{
|
|
resetDevice();
|
|
selected_session = session;
|
|
requestData();
|
|
}
|
|
|
|
void CMS50F37Loader::resetDevice()
|
|
{
|
|
qDebug() << "cms50f37 - Resetting oximeter";
|
|
sendCommand(COMMAND_CMS50_HELLO1);
|
|
QThread::msleep(100);
|
|
QApplication::processEvents();
|
|
sendCommand(COMMAND_CMS50_HELLO2);
|
|
QThread::msleep(100);
|
|
QApplication::processEvents();
|
|
}
|
|
|
|
void CMS50F37Loader::requestData()
|
|
{
|
|
qDebug() << "cms50f37 - Requesting session data";
|
|
sendCommand(COMMAND_GET_SESSION_DATA, selected_session);
|
|
}
|
|
|
|
void CMS50F37Loader::killTimers()
|
|
{
|
|
if (resetTimer.isActive()) resetTimer.stop();
|
|
if (startTimer.isActive()) startTimer.stop();
|
|
}
|
|
|
|
void CMS50F37Loader::startImportTimeout()
|
|
{
|
|
}
|
|
|
|
void CMS50F37Loader::resetImportTimeout()
|
|
{
|
|
if (finished_import) {
|
|
return;
|
|
}
|
|
|
|
if (imp_callbacks != cb_reset) {
|
|
// Still receiving data.. reset timer
|
|
qDebug() << "cms50f37 - Still receiving data in resetImportTimeout()" << imp_callbacks << cb_reset;
|
|
if (resetTimer.isActive())
|
|
resetTimer.stop();
|
|
|
|
if (!finished_import)
|
|
resetTimer.singleShot(2000, this, SLOT(resetImportTimeout()));
|
|
} else {
|
|
qDebug() << "cms50f37 - Oximeter device stopped transmitting.. Transfer complete";
|
|
qDebug() << "cms50f37 - Import packet count: " << importCount;
|
|
// We were importing, but now are done
|
|
if (!finished_import && (started_import && started_reading)) {
|
|
qDebug() << "cms50f37 - Switching CMS50F37 back to live mode and finalizing import";
|
|
// Turn back on live streaming so the end of capture can be dealt with
|
|
|
|
|
|
killTimers();
|
|
|
|
finished_import = true;
|
|
m_streaming = false;
|
|
m_importing = false;
|
|
started_reading = false;
|
|
started_import = false;
|
|
|
|
emit importComplete(this);
|
|
|
|
m_status = NEUTRAL;
|
|
|
|
shutdownPorts();
|
|
|
|
return;
|
|
}
|
|
qDebug() << "cms50f37 - Should CMS50F37 resetImportTimeout reach here?";
|
|
// else what???
|
|
}
|
|
cb_reset = imp_callbacks;
|
|
}
|
|
|
|
void CMS50F37Loader::shutdownPorts()
|
|
{
|
|
closeDevice();
|
|
}
|
|
|
|
|
|
|
|
|
|
bool CMS50F37Loader::readSpoRFile(const QString & path)
|
|
{
|
|
QFile file(path);
|
|
if (!file.exists()) {
|
|
qWarning() << "cms50f37 - Can't find the oximeter file: " << path;
|
|
QMessageBox::warning(nullptr, STR_MessageBox_Error, "<h2>"+tr("Could not find the oximeter file:")+"<br/><br/>"+path+"</h2>");
|
|
return false;
|
|
}
|
|
|
|
if (!file.open(QFile::ReadOnly)) {
|
|
qWarning() << "cms50f37 - Can't open the oximeter file: " << path;
|
|
QMessageBox::warning(nullptr, STR_MessageBox_Error, "<h2>"+tr("Could not open the oximeter file:")+"<br/><br/>"+path+"</h2>");
|
|
return false;
|
|
}
|
|
|
|
bool spo2header = false;
|
|
QString ext = path.section('.', -1);
|
|
qDebug() << "cms50f37 - Oximeter file extention is " << ext;
|
|
if (ext.compare("spo2",Qt::CaseInsensitive) == 0) {
|
|
spo2header = true;
|
|
qDebug() << "cms50f37 - Oximeter file looks like an SpO2 type" ;
|
|
}
|
|
|
|
QByteArray data;
|
|
|
|
qint64 filesize = file.size();
|
|
data = file.readAll();
|
|
QDataStream in(data);
|
|
in.setByteOrder(QDataStream::LittleEndian);
|
|
quint16 pos;
|
|
in >> pos;
|
|
|
|
in.skipRawData(pos - 2);
|
|
|
|
//long size = data.size();
|
|
int bytes_per_record = 2;
|
|
|
|
if (!spo2header) {
|
|
// next is 0x0002
|
|
// followed by 16bit duration in seconds
|
|
|
|
// Read date and time (it's a 16bit charset)
|
|
|
|
char dchr[20];
|
|
int j = 0;
|
|
for (int i = 0; i < 18 * 2; i += 2) {
|
|
dchr[j++] = data.at(8 + i);
|
|
}
|
|
|
|
dchr[j] = 0;
|
|
if (dchr[0]) {
|
|
QString dstr(dchr);
|
|
// Ensure date is correct first to ensure DST is handled correctly
|
|
QDate date = QDate::fromString(dstr.left(8),"MM/dd/yy");
|
|
QTime time = QTime::fromString(dstr.right(8),"HH:mm:ss");
|
|
if (date.year() < 2000) {
|
|
date = date.addYears(100);
|
|
}
|
|
m_startTime = QDateTime(date, time);
|
|
} else { // this should probaly find the most recent SH data day
|
|
m_startTime = QDateTime(QDate::currentDate(), QTime(0,0,0)); // make it today at midnight
|
|
cms50dplus = true;
|
|
}
|
|
} else { // it is an spo2header
|
|
|
|
quint32 samples = 0; // number of samples
|
|
|
|
quint32 year, month, day;
|
|
quint32 hour, minute, second;
|
|
|
|
if (data.at(pos) != 1) {
|
|
qWarning() << "cms50f37 - .spo2 file" << path << "might be a different";
|
|
}
|
|
|
|
// Unknown cruft header...
|
|
in.skipRawData(200);
|
|
|
|
in >> year >> month >> day;
|
|
in >> hour >> minute >> second;
|
|
|
|
if ( year+month+day == 0 ) {
|
|
m_startTime = QDateTime(QDate::currentDate(), QTime(0,0,0));
|
|
cms50dplus = true;
|
|
}
|
|
else
|
|
m_startTime = QDateTime(QDate(year, month, day), QTime(hour, minute, second));
|
|
|
|
// ignoring it for now
|
|
pos += 0x1c + 200;
|
|
|
|
in >> samples;
|
|
|
|
int remainder = filesize - pos;
|
|
|
|
bytes_per_record = remainder / samples;
|
|
qDebug() << "cms50f37 - " << samples << "samples of" << bytes_per_record << "bytes each";
|
|
|
|
// CMS50I .spo2 data have 4 digits, a 16bit, followed by spo2 then pulse
|
|
|
|
}
|
|
|
|
oxirec = new QVector<OxiRecord>;
|
|
oxisessions[m_startTime] = oxirec;
|
|
|
|
unsigned char o2, pr;
|
|
quint16 un;
|
|
|
|
// Read all Pulse and SPO2 data
|
|
do {
|
|
if (bytes_per_record > 2) {
|
|
in >> un;
|
|
}
|
|
in >> o2;
|
|
in >> pr;
|
|
|
|
if ((o2 == 0x7f) && (pr == 0xff)) {
|
|
o2 = pr = 0;
|
|
un = 0;
|
|
}
|
|
|
|
if (spo2header) {
|
|
oxirec->append(OxiRecord(pr, o2));
|
|
} else {
|
|
oxirec->append(OxiRecord(o2, pr));
|
|
}
|
|
} while (!in.atEnd());
|
|
|
|
|
|
// for (int i = pos; i < size - 2;) {
|
|
// o2 = (unsigned char)(data.at(i + 1));
|
|
// pr = (unsigned char)(data.at(i + 0));
|
|
// oxirec->append(OxiRecord(pr, o2));
|
|
// i += 2;
|
|
// }
|
|
|
|
// processing gets done later
|
|
return true;
|
|
}
|
|
|
|
void CMS50F37Loader::process()
|
|
{
|
|
// Just clean up any extra crap before oximeterimport parses the oxirecords..
|
|
return;
|
|
// if (!oxirec)
|
|
// return;
|
|
// int size=oxirec->size();
|
|
// if (size<10)
|
|
// return;
|
|
|
|
}
|
|
|
|
|
|
|
|
static bool cms50f37_initialized = false;
|
|
|
|
void CMS50F37Loader::Register()
|
|
{
|
|
if (cms50f37_initialized) { return; }
|
|
|
|
qDebug() << "Registering CMS50F37Loader";
|
|
RegisterLoader(new CMS50F37Loader());
|
|
cms50f37_initialized = true;
|
|
}
|
|
|